a Department of Biomedical Engineering, Tufts University, Boston, MA b Department of Biomedical Engineering, The University of Texas at Austin, Austin, TX c Department of Physics, Southwestern University, Georgetown, TX d Department of Bioengineering, University of Pittsburgh, Pittsburgh, PA e Vascular Medicine Institute, University of Pittsburgh Medical Center, Pittsburgh, PA 15213 f Department of Medicine, University of Pittsburgh, Pittsburgh, PA 15213
The complete text follows below, reproduced from the open access article under CC BY 4.0. Figures are in the PDF and in the illustrated overview.
The Allstruder is a modular syringe pump and extruder system engineered to combine high performance, broad compatibility, and exceptional ease of use. It is a compelling alternative to existing material extrusion tools across bioprinting, embedded 3D printing, food printing, ceramics, and multi-material applications. Validated across multiple laboratories and use cases, the Allstruder has demonstrated reliable performance and adaptability with a wide range of materials, syringe sizes, and 3D printer platforms. Designed specifically for additive manufacturing, it is configurable for nearly any extrusion-based 3D printer, supports syringes from 0.1 to 100 mL, and relies exclusively on standard, off-the-shelf components. Its architecture emphasizes adaptability through modular mounts and interchangeable syringe adapters, allowing rapid reconfiguration for different tools, materials, and workflows. With a simple, durable, and modifiable design, the Allstruder supports both novice users and experienced researchers, providing a robust platform for reproducible and high-resolution syringe-based extrusion in scientific, educational, and industrial environments.
Keywords: Bioprint, 3D print, Additive manufacturing, Syringe pump, Extruder, Printer modification
| Hardware name | The Allstruder Syringe Pump Extruder |
|---|---|
| Subject area | Engineering and materials science |
| Hardware type | Other: Additive manufacturing of fluids and pastes |
| Closest commercial analog | CellInk BioX |
| Open source license | CERN OHL v2 |
| Cost of hardware | 51 USD |
| Source file repository | doi.org/10.5281/zenodo.15670349 |
| Secondary repository | github.com/hintonthomas/Allstruder |
The increasing demand for high performance 3D printing across bioprinting, food printing, ceramics, electronics, and soft robotics has exposed the limitations of existing extrusion technologies. Commercial systems typically fall into two categories: volumetrically accurate syringe pumps that lack responsiveness, or pressure-based syringe extruders that are fast to develop pressure but imprecise [1]. Both are expensive, difficult to customize, and rarely compatible with the open, modular workflows used in many research labs and educational settings. This gap poses a technical barrier that drives many labs to reinvent hardware to meet the same unmet need, failing to account for reproducibility and accessibility between users. A shared, high-performance, and interoperable standard for a syringe extruder is how open-source innovation could unite various additive manufacturing communities.
Despite these limitations, no widely adopted open-source alternative has emerged. The rise of accessible desktop thermoplastic 3D printers has inspired a wave of DIY and open-source fluid extrusion systems, many of which are creative and resourceful [2–24]. However, most are tailored for specific needs, materials, or machines, and few are engineered to generalize across use cases, users, and environments. The result is a cluttered design space with dozens of extruders that partially solve a problem but rarely scale beyond a single lab or community. As a result, labs are frequently forced to tailor their workflows to the constraints of a specific extruder, locking themselves into siloed methodologies that limit reproducibility, interoperability, and shared progress. Such fragmentation can elicit reinvention of existing solutions and slow research progress; for example, a graduate student should not devote significant time to developing an in-house extruder for their desired application. To address this common pitfall of lab-locked innovation, an open-source design should excel across performance, accessibility, and adaptability.
While most DIY syringe extruders have fallen short of general-purpose adoption, several designs stand out as foundational and have meaningfully advanced the field:
These designs and many others helped demonstrate the value of open-source hardware in advancing syringe-based extrusion. However, no single extruder has yet achieved the necessary balance of performance, adaptability, and accessibility to unify researchers across disciplines and reduce the fragmentation of the field of fluids additive manufacturing.
The Allstruder was developed not just to improve performance, but to serve as an open, generalizable, robust, and user-friendly platform for syringe-based extrusion and additive manufacturing. It was designed in direct response to the proliferation of incompatible extruder projects, with the goal of supporting reproducible research, community-driven innovation, and standardization. Key features include:
The Allstruder has already been tested across multiple labs and application areas, including printing of hydrogels, pastes, and viscous fluids prone to bubble retention or non-Newtonian flow. Its modular design allows it to be easily mounted on diverse printer architectures and reconfigured for specialized use cases.
Beyond its mechanical design, the Allstruder represents an effort to standardize a common foundation for the next generation of syringe-based printing. By offering an extrusion system that is broadly compatible, reproducible, and open to modification, we hope to reduce redundancy in future design efforts and encourage community collaboration rather than reinvention.
Our aim is to consolidate lessons from past tools within a single platform that is ready for adoption and extension. The Allstruder is offered as a freely accessible, adaptable, high-performance solution that invites participation from students, educators, researchers, and advanced users alike, and may serve as a unifying base for reproducible manufacturing of complex fluids.
The actuator (Fig. 1B) forms the backbone of the Allstruder and provides the structural framework for mounting to a 3D printer. It consists of a 3D printed frame that houses a leadscrew-driven pusher block, which is guided along a precision linear rail. The leadscrew is supported by ball bearings on both ends, ensuring smooth, stable motion. The pusher block is supported by the linear rail truck and features a keyslot and through-holes for attaching the syringe pusher adapter. This design enables controlled depression and retraction of the syringe plunger. The actuator comes in two size variants: a regular version with a 66 mm stroke and an extended version with a 116 mm stroke, accommodating different syringe volumes and motion requirements.
The transmission (Fig. 1C) is housed within a 3D printed motor mount and consists of a timing belt connecting the motor output to the actuator's lead screw. The motor mounts to the motor mount using slotted holes, which allow it to slide backward for tensioning the timing belt. This indirect drive setup, when used with a quality, fiber-backed timing belt such as 2GT, combines straightforward tensioning and backlash reduction with smooth torque transfer. The motor mount also includes two front-facing bosses with square nut slots for mounting the syringe holder adapter. The transmission is currently available in two versions optimized for different NEMA motor sizes: NEMA 17 (requires a 180 mm belt) and NEMA 14 (requires a 120 mm belt) and is easily configurable to other motor mounting patterns and sizes.
The motor (Fig. 1D) serves as the driving force behind the Allstruder. Currently, NEMA 17 and NEMA 14 stepper motors are supported, but any motor that fits within the space underneath the motor mount can be made to work.
The syringe interface (Fig. 1E) allows the Allstruder to handle a wide range of plunger-driven syringe sizes, including reusable gastight Hamilton syringes from 0.1 to 25 mL and disposable plastic syringes from 3 to 100 mL. Each syringe is connected using a pair of 3D printed adapters: the pusher attaches to the syringe plunger and mounts to the pusher block, while the holder attaches to the syringe barrel and secures to the motor mount. This modular adapter system enables easy swapping of syringes and maintains alignment during extrusion. Each adapter set is designed to position the syringe as close as possible to the actuator assembly, which serves as the primary mechanical reference. This mechanical optimization minimizes deflection in the pusher, motor mount bosses, and adapter interfaces, thereby reducing positional inaccuracies during extrusion.
In developing the Allstruder, we combined our experience designing the published open-source extruders shown in Fig. 2A with insights gained from working with commercial bioprinting systems from brands such as Allevi (3D Systems), CellInk (BICO), RegenHU, and Puredyne (Viscotec). We also gathered insight from several unpublished or proprietary extruder systems, some of which are specialized for manufacturing, high-pressure, high-volume, or high-speed applications. The goal was to integrate the most effective features from across the ecosystem into a design that maximized utility without introducing unnecessary complexity or restricting users.
To better understand the limitations of existing DIY syringe extruders and identify unmet needs, we developed a five-metric framework to evaluate their utility and guide Allstruder design (Fig. 2B). We believe affordability, performance, simplicity, versatility, and design for 3D printing capture important design principles for open usability and reproducibility in open-source syringe pump extruders.
Together, these five metrics offer a structured way to compare syringe pump designs and assess their usefulness in real-world printing contexts. To apply this framework consistently, we decomposed each of the five metrics into specific, measurable sub-criteria and scored every extruder against them. The complete rubric, sub-criteria, and per-device scores are provided in the accompanying spreadsheet, so that each score can be traced to its basis and the same rubric applied to extruders not considered here. The resulting scores are summarized in Fig. 2B-C and are best understood not as rankings but as a map of deliberate design trade-offs.
These scores also make clear that each prior extruder represents a deliberate optimization for a particular corner of the design space rather than a failure to reach an ideal, and the differences in Fig. 2B-C largely reflect those choices and the era in which each tool was made. The Replistruder 3 was produced when no comparable open syringe extruder existed, and it prioritized affordability, printability, and performance on low-cost hardware. In expert hands it produced some of the best embedded bioprinting of its time, but it required considerable expertise and was not intended to be simple to interpret, which is why it scores well on affordability and performance but low on simplicity. The Replistruder 4 responded to those limitations with a structural change, introducing off-the-shelf metallic elements into an otherwise 3D-printed mechanism to trade fully printed construction for the stiffness of aluminum and steel. This hybrid approach preserved much of the accessibility and prototypability of the series while delivering a substantial gain in performance, and it was supported by thorough, widely circulated documentation. The Replistruder 5 carried that trajectory to its conclusion as a lightweight, compact, high-precision extruder in which nearly all of the cost is invested in a premium leadscrew, rail, and motor. It ranks highest in mechanical performance because it is optimized entirely around glass syringes and a narrow, high-quality configuration for cartridge-style, multi-toolhead systems, and this same specialization is why it scores lower on versatility and affordability, by intent rather than by oversight. The Enderstruder occupies the opposite corner, forgoing optimization and breadth to maximize simplicity and accessibility for a single popular low-cost printer, while the Large Volume Extruder accepts the resolution ceiling imposed by a 50 mL syringe, which offsets the benefit of better hardware, in exchange for the large volumes and low-cost materials such as clays and alginates that made it broadly useful.
Taken together, these designs show that a high score in one dimension is typically obtained at the expense of another, and that no prior tool occupies the center of the space. The Allstruder was developed against a different objective, remaining broadly capable across all five metrics at once and targeting the simplicity and versatility we identified as the primary gaps left by existing DIY designs. On parts designed for 3D printing it matches the Replistruder 5, as both are current designs shaped by extensive experience and focused on reliability and accessibility, and the core Allstruder components have been refined through more than 85 iterations, most aimed at improving success rates during printed part production on standard desktop FFF machines. At $51 per extruder, the Allstruder ranks just behind the Replistruder 3 in affordability while avoiding that design's performance compromises. It does not surpass a purpose-built tool within that tool's own niche, and it will not out-perform a Replistruder 5 paired with a glass syringe on a high-quality motion system, but it delivers strong performance while remaining easier to use and more broadly compatible, which makes it the more practical option across most users and applications. This versatility follows directly from the modular, reconfigurable drive detailed in the Configurability, Mounts, and Adapters section and illustrated in Fig. 5, which supports the syringes shown in Fig. 2E and the use cases shown in Fig. 2F, and allows users to change syringes, substitute stronger components, or extend the mechanical stroke over the course of a biofabrication campaign, from conceptualization through to application and publication. Critically, the mechanical performance retained across all these configurations is governed by proper assembly of the Allstruder components (Figs. 6–11), and is governed by the rigidity of the drive mechanism, quantified in Fig. 13, rather than by any single syringe or printer, so this breadth does not come at the expense of the determinant of extrusion fidelity.
To maximize simplicity, the Allstruder was designed with as few components as possible. The Allstruder also needed to accommodate a wide range of use cases, necessitating a modular and adaptable construction. As illustrated in Fig. 3A, users can easily swap in different motors, linear rails, lead screws, and belted transmissions to configure the system for various mechanical requirements. To support different syringe strokes, the actuator is adjustable in length, and the transmission assembly can be scaled to account for varying motor sizes, belt lengths, and pulley dimensions. The frame, which houses the actuator, is available in standard and extended sizes, offering stroke lengths of 66 mm and 116 mm, respectively. Similarly, the motor mount comes in two sizes: one designed for NEMA 17 motors paired with a 180 mm belt, and a more compact version for NEMA 14 motors using a 120 mm belt.
Allstruder mechanical components were selected for their availability, durability, and performance, making them both reliable and easy to replace. The pusher assembly converts the rotary motion of the lead screw into linear displacement of the syringe plunger. To enhance stiffness, long bolts run the length of the pusher, applying preload across 3D printed parts. The lead screw nut is secured to the pusher block using self-tapping screws, which are easier to install than heat-set inserts and offer stronger, more reliable fastening than conventional bolts threaded into plastic. The surrounding frame fully encloses the pusher, maximizing mechanical support.
Since the pusher is subject to bending and compression forces and must be partially disassembled during syringe adapter changes, proper realignment during reassembly is essential. To address this, the front face of the pusher block includes a keyed slot (Fig. 3E) that mates with a boss on the rear of each syringe adapter. This feature ensures correct assembly every time a syringe adapter is swapped.
To ensure the Allstruder would perform reliably in high-speed and multi-material 3D printing applications, we paid close attention to its physical footprint and dynamic behavior as a moving payload. Much like a high-performance racing vehicle with a wide wheelbase, the Allstruder uses four widely spaced mounting bolts (Fig. 4A) to secure itself firmly to a printer's gantry, enhancing stability during rapid movements. Its low center of mass (Fig. 4B) further minimizes vibration and wobble caused by inertia, particularly during sharp directional changes at speed.
While compact extruders like the Replistruder 5 are optimized to conserve lateral space in high performance, multi-extruder systems, the Allstruder prioritizes versatility [30]. To support high-torque actuation and broad hardware compatibility, the Allstruder accommodates motors up to NEMA 17 size, and since the motor is the widest component, it determines the overall Allstruder width of approximately 44 mm (Fig. 4C). For multi-material configurations, the Allstruder supports both direct drive and Bowden-style setups. Multiple Allstruders can be mounted close together to minimize space or further apart, depending on the needs of the user. A potential case for mounting Allstruders further apart might be for printing multiple materials into standard petri dishes, such that a secondary, standby Allstruder's needle does not collide with the dish while another primary Allstruder is extruding inside it. When configured in a coaxial extrusion system (Fig. 4D), the Allstruder can operate either as a gantry-mounted extruder for small-volume inks or as a frame-mounted pump feeding larger volumes through tubing to the printhead. Many of the printer-specific mounts displayed in Fig. 5C incorporate mounting points for multiple Allstruders to natively support multi-material setups.
To clearly illustrate how the Allstruder can be configured for specific applications, three example setups are shown in Fig. 5A and described in detail in Fig. 5B:
The Allstruder platform offers modularity comparable to a LEGO set. Users can mix and match components (as shown in Fig. 5C, and beyond) to create a setup tailored to their specific needs. The growing ecosystem currently includes: 2 frame options, 2 motor mounts, 2 pusher blocks, 12 syringe adapters, 1 Bowden nozzle adapter, 1 syringe heater (in development), and over 10 printer adapters. In general, users are encouraged to begin component selection with the Frame and Lead Screw and proceed clockwise through the options, concluding with the Printer Mount/Adapter.
The Allstruder has been and can be utilized in a variety of ways in and outside the original user community. Below are a few examples of unique or novel use cases:
All design files are available as STP and STL under CERN OHL v2 at doi.org/10.5281/zenodo.15670349.
| Design file | Description |
|---|---|
| Frame | The largest single part. Houses the linear actuator components including the lead screw, linear rail, and pusher assembly. |
| Frame_Extended | A 50 mm taller variant of the frame, intended for use with 200 mm leadscrews and 150 mm linear rails. |
| Pusher_Block | The mobile element in the actuator, connecting the lead screw nut and the linear rail truck to actuate the syringe pusher adapter. |
| Pusher_Block_Anti-Backlash_Top | First half of a pusher block variant designed to accommodate anti-backlash nuts, for extreme precision and low thrust applications with microliter volume syringes. |
| Pusher_Block_Anti-Backlash_Bottom | Second half of the anti-backlash pusher block variant. |
| Knob | A knurled attachment enabling manual actuation of the lead screw. |
| Motor_Mount | Housing for the transmission. Serves as the mounting point for syringe holder adapters and connects the actuator to any NEMA 17 motor using a 180 mm belt. |
| Motor_Mount_NEMA_14_120mm_Belt | A shorter motor mount variant connecting any NEMA 14 motor to the actuator using a 120 mm belt. |
| Syringe_Pusher | Top half of a syringe adapter. Actuates the plunger and connects to the pusher block. One of many adapters covering the syringes in Fig. 2E. |
| Syringe_Holder | Bottom half of a syringe adapter. Holds the barrel stationary and connects to the motor mount. |
| Needle_Mount_Bowden_Extruder | Allows a needle with tubing to connect in place of an Allstruder on a printer's extruder gantry, using the same printer mount. Intended for stainless steel and coaxial needles. |
| Allstruder_Mount_Printer | Printer-specific mounting adapters. All use the same 35 × 50 mm four-bolt pattern. |
| Component | Qty | SKU | Per SKU | USD/Allstruder |
|---|---|---|---|---|
| 100 mm MGN9C Linear Rail Kit | 1 | B07WNXP2K5 | $7.89 | $7.89 |
| 150 mm Tr8X8 Leadscrew, Nut | 1 | B07C8P1DWX | $12.99 | $12.99 |
| 608 ZZ/RS Bearings | 2 | B07R7PR72H | $7.99 | $1.60 |
| 20 T 2GT Pulley, 5 mm bore | 1 | B077GNZK3J | $6.99 | $1.40 |
| 40 T 2GT Pulley, 8 mm bore | 1 | B09JSMRZH9 | $4.99 | $2.50 |
| NEMA 17 Motor, 63.74 oz-in | 1 | B07LF898KN | $11.99 | $11.99 |
| 180 mm 2GT Belt | 1 | B014QLQBQ0 | $9.99 | $1.00 |
| 8 mm Shaft Collar | 1 | B0BZ7C2X2W | $8.99 | $0.75 |
| Thin M8 Washer | 1 | B0C3TV2QDX | $5.89 | $0.29 |
| Thin Square M3 Nuts | 1 | B0D6SPHN8G | $11.49 | $0.23 |
| 8 mm M4 Bolt | 1 | B0BHQZHSLQ | $8.38 | $0.08 |
| 10 mm M3 Bolts | 15 | B08R3GVJVC | $8.26 | $1.24 |
| 20 mm M3 Bolts | 2 | B089KT7ZNW | $8.96 | $0.18 |
| 40 mm M3 Bolts | 2 | B0BGM73YG8 | $8.22 | $0.16 |
| 45 mm M3 Bolts | 4 | B089KRYJ9Q | $8.42 | $0.67 |
| 8 mm M3 Self Tapping Screw | 4 | B09CPHWX7P | $10.99 | $0.07 |
| PETG Plastic Filament | ~200 g | B07PGYHYV8 | $15.99 | $3.20 |
The Allstruder is built using hardware listed in the Bill of Materials along with 3D printed components, and it can be divided into four main sections. Before assembly, all printable parts and required hardware should be collected. Assembly primarily occurs in a single plane. Once assembled, the Allstruder can function as a standalone syringe pump or as an extruder mounted to a printer.
A build video explaining the assembly (Figs. 6–11) is the recommended method for following along and assembling the Allstruder, and it can be accessed through the supplemental information for this paper and alternatively through the Allstruder GitHub. To build the Allstruder, a user will first need to gather the hardware listed in the BOM and print the 3D printable components. The printable components of the Allstruder were designed to be printed with PETG, and success has also been achieved with PCTG and some photopolymer resins such as Formlabs Clear Resins. PLA is not recommended for long-term use because of its tendency to creep under load, and ABS, ASA, and PC are only recommended with a 3D printer that can mitigate part deformation during printing. Allstruder part files are available as STL, 3MF, and STP, and each file is in the correct orientation for FFF printing. Project files with the Allstruder parts pre-oriented with appropriate settings for PETG on a Prusa MK4 are available through the supplemental information for this paper and alternatively through the Allstruder GitHub.
The Allstruder operates similarly to other DIY syringe pump extruders. The general procedure for setup and use is as follows.
After assembling the Allstruder, the next step is checking that the assembly operates as expected. Rotation of the leadscrew should be possible using minimal manual force, and there should be no axial slop in the leadscrew itself. If rotation of the leadscrew requires a great deal of torque, check that the leadscrew nut is correctly installed in the pusher. If the leadscrew can be moved along its axis, repeat step 16 of the assembly, ensuring that the leadscrew is pushed toward the collar nut while it is being installed around the top of the leadscrew, so as to eliminate any axial slop. The leadscrew should be unable to wiggle but easy to rotate. If repeating step 16 multiple times does not resolve axial slop, it may be necessary to rotate the collar a quarter turn before re-tightening it onto the leadscrew to prevent the collar's grub screw, if it has one, from pushing the collar up. Other potential problems to watch for include loose pulleys. Double check that no pulley can move along the leadscrew or motor shaft. If a pulley is loose, tighten the pulley screws until the pulley is securely attached.
Following mechanical function verification, it is time to install the Allstruder on a 3D printer. Most desktop 3D printers using open-source firmware such as Marlin, RepRapFirmware, or Klipper are compatible with the Allstruder due to their use of standard stepper motors. Closed-firmware systems are not recommended, though their motion hardware can often be repurposed by replacing the controller with an open-source board, such as a Duet. Mounting an Allstruder to a suitable printer requires fabricating and installing a compatible mount. For many common printer models, we have provided 3D-printable mounts that are available in the supplementary materials of this paper or on the Allstruder GitHub repository. For printers without an existing mount, users may need to remove the original thermoplastic extruder, carefully measure the extruder gantry, and prototype a custom mount before installing it in place of the original component.
Once mounted, the Allstruder must be electrically connected to the printer or motor controller. This usually involves attaching a 4-wire cable from the Allstruder's stepper motor to the controller, typically a standard 3D printer motherboard. Soldering is rare. After the connection is made, the motor controller must be properly configured to operate the Allstruder. For firmware on most 3D printers, the most critical setting is the steps/mm value used for extrusion. For a standard Allstruder equipped with a TR 8x8 lead screw, the steps/mm can be calculated as follows.
For Klipper firmware-based 3D printers, the equivalent value used to control the extruder stepper movement is rotation distance and is calculated as follows.
Ideally, the syringe is loaded with material in a manner that avoids trapping or generating bubbles, since bubbles always negatively impact the predictability of extrusion. There are a few places where bubbles can become trapped within the system: within the material, in a dead volume space such as that found within a needle hub, and against a surface such as a black rubber plunger. Alternatively, bubbles can be generated by ink materials that were previously mixed and refrigerated but have been left out at ambient temperature. Warming causes dissolved gasses to nucleate within the syringe and needle, and as such, printing materials should be degassed only after warming to room temperature.
To load a syringe and needle without bubbles, there are many methods, and the following steps should cover most materials in most plastic syringes:
There are many effective ways to set up G-code for printers using one or more experimental extruders, making it impossible to cover every possible configuration here. Likewise, G-code can be structured in numerous ways depending on the material and printing approach, and many different software tools can generate valid results. As one example, the following settings demonstrate how PrusaSlicer can be used to generate G-code for printing a single material with an Allstruder-equipped printer configured to act purely as a G-code interpreter. This process does not require specialized pre- and post-processing scripts such as homing and calibration, which are typically used in thermoplastic printing. For users running Klipper firmware, executing Allstruder-specific G-code may require disabling or bypassing macros such as START_PRINT and END_PRINT, which are normally associated with thermoplastic workflows. The extremely simplified example below was developed to enable direct ink writing of mayonnaise, that is, printing mayonnaise in air using a 5 mL plastic syringe with a 27 gauge needle on a Creality K1 SE fitted with an Allstruder. The corresponding PrusaSlicer profile is available in the Allstruder GitHub repository.
| Layer height | 0.1 |
|---|---|
| Extrusion width | 0.25 |
| Number of perimeters | 2 |
| Infill | 100% |
| Extrusion speed | 23 mm/s |
| Travel speed | 150 mm/s |
| Retraction / deretraction distance | 0.6 mm |
| Retraction / deretraction speed | 0.5 mm/s |
| Z-hop during travel | 0.25 mm |
| Start G-code | SET_KINEMATIC_POSITION X = 110 Y = 110 Z = 0 E = 0 |
| End G-code | CANCEL_PRINT |
Load the prepared syringe into the Allstruder. The Allstruder is designed for front-loading, allowing the syringe to be inserted with a simple press fit. If the syringe feels loose, the adapter may need to be modified, or thin layers of tape can be applied to the syringe barrel to create a snug, stable fit.
Next, prepare the print container, with or without support material. For printing in air, secure the surface that will serve as the print base. For embedded or FRESH printing, secure the container holding the support material and ensure the needle can enter the bath without obstruction, verifying that the syringe and needle hub will not collide with the sides of the container during movement.
Before starting the print, confirm that the Allstruder can extrude properly. Clogging or lack of extrusion can result from dried material on the needle tip, a disconnected motor cable, or solidified material inside the syringe. A simple test is to command a small extrusion via the printer's control interface and visually confirm continuous material flow from the needle. Then wipe the needle clean before proceeding.
Finally, position the needle or nozzle tip at the correct starting location. For printing in air, align the needle tip just above the print surface, for example the base of a petri dish. For embedded or FRESH printing, position the needle at the desired depth within the support material, ensuring it is fully submerged and clear of nearby walls or obstructions.
Once the print is initiated, there are several factors to monitor during operation. Many printer firmwares and default slicing profiles include start and end scripts designed for conventional FFF/FDM printing workflows. These profiles handle homing, heating, and cooldown sequences automatically, and can be defined for specific materials. When using an Allstruder, these scripts can interfere with the desired behavior. Unintended motions, fans starting, or automatic heating of the thermoplastic extruder or build plate are clear signs that such scripts are still active and should be disabled. During longer prints, particularly in dry environments, users should also be aware of potential dehydration of either the printed material itself or, in the case of embedded printing, the support bath. Air bubbles within the syringe or needle can lead to intermittent flow or cause retraction and priming moves to become unresponsive. In addition, small accumulations of material around the needle tip may interfere with deposition if they contact previously extruded material.
The Allstruder was tested across a diverse network of labs, each with unique materials, methods, printers, and user experience levels. By distributing the design and build instructions broadly and collecting iterative feedback, we identified and resolved adoption-limiting issues, resulting in a robust, user-friendly tool capable of reproducible printing across a wide range of settings. This process validated the Allstruder's performance with hydrogels, pastes, and other viscous or non-Newtonian materials, while also confirming its compatibility with varied printer architectures and syringe extrusion workflows.
Integrating the Allstruder into each research setup required understanding the specific needs of the user. These needs can vary but normally include their printer model, printing material, target object, and syringe requirements. For example, the Printess, a low-cost, open-source, small-format bioprinter designed by the Skylar-Scott lab, is a highly modular extrusion-based 3D printing platform that was made compatible with the Allstruder by designing a custom mount tailored to its dimensions [24]. This adaptation enabled the Printess to benefit from the Allstruder's high torque and large-volume extrusion capabilities, while continuing to provide affordability. Across the remainder of Fig. 12A, a wide variety of printers have been adapted to use the Allstruder, ranging from affordable Creality Ender 3 V2s to a CellInk INKREDIBLE. These integrations have led to the creation of numerous novel, largely unpublished materials and printer configurations. Many researchers have contributed to the development of the Allstruder by offering critical feedback on its functionality and helping to expand the library of components, including new printer mounts and syringe adapters.
The Allstruder's actuator was mechanically tested to understand how rigid it is under loading. A Mitutoyo dial indicator was used to measure deflection of the pusher assembly when an inserted 5 mL plastic syringe was pressurized to 30 PSI or 207 kPa above ambient, which exceeds the pressure necessary to print most bioinks. For this characterization, the Allstruder's 3D-printed components were fabricated, using settings found in the PrusaSlicer project available on the Allstruder GitHub repository, with PETG filament, 2–6 perimeters, 20% infill, and 5–7 top and bottom layers. The result was a maximum deflection of 20 µm at a calculated force of 24 N. While this level of error is minimal and compounded by the compliance of plastic syringes, which introduce additional material flow under pressure, it reflects a conservative, worst-case scenario in commercial bioprinters using air-driven extrusion at maximum operating pressure. Error volume will depend on the syringe used in any given application, with larger plastic syringes having the greatest. For users requiring greater accuracy at high pressures, replacing the plastic syringe with a smaller or glass alternative would reduce deformation and improve volumetric precision.
To complement the device-level deflection measurement, we quantified the output of a single complete configuration under conditions representative of ordinary use. Using the Creality K1 SE 3D printer, a 1 mL Hamilton Glass syringe, 23 mg/mL Type 1 Collagen (LifeInk 240), and a 30 gauge by 0.5 in. needle, we printed a resolution test pattern with two distinct and common filament spacings of 1000 and 500 µm, and measured extruded feature width and dimensional error. Printed filament width was shown to be a mean of 141.8 µm ± 15.61 µm with a small 6% error from the 150 µm needle used for printing. The overall configuration produced results with good repeatability, and the greatest error, the y axis dimension of overall print width at 4.6% error, likely stems from the host printer's mechanics and not the Allstruder. Center-to-center spacing of each spacing condition was below 3% error for each designed spacing. Spacing between each filament, which is set by subtracting the filament width of 150 µm from the designed center-to-center spacing, which equates to a spacing target of 350 µm and 850 µm for 500 µm and 1000 µm respectively, showed similar small print errors of less than 4%. As with any extrusion result, these values reflect the full configuration rather than the extruder alone. The choice of syringe is particularly influential, since barrel compliance, stiction, and seal friction vary substantially across the 0.1 to 100 mL range the Allstruder supports, and a smaller, stiffer glass syringe would be expected to yield tighter tolerances than a large plastic one. Fig. 14 therefore serves as a worked example of the Allstruder's achievable output quality rather than a bound on it.
Together, these measurements characterize the Allstruder at two distinct levels. The deflection test in Fig. 13 quantifies the rigidity of the drive mechanism itself, a property intrinsic to the extruder that bounds achievable fidelity independently of the host printer or syringe. The resolution measurement in Fig. 14 instead quantifies the output of one complete configuration, comprising a specific host printer, syringe, material, and needle. Metrics like the latter are inherently configuration-dependent. Because the drive is a deterministic leadscrew-and-rail mechanism, the accuracy and repeatability of a finished print are governed largely by the compliance, stiction, and seal friction of the chosen syringe, by the material, and by the motion system of the host machine, rather than by the extruder in isolation. The values in Fig. 14 should therefore be read as a representative example for the stated configuration rather than as a general specification. Taken as a whole, this is the balance the Allstruder was built to strike: a mechanism rigid enough that fidelity is limited by the user's choice of syringe and printer rather than by the extruder, offered as one open and inexpensive platform that adapts to the material, the machine, and the application at hand. It is in that adaptability, demonstrated across many laboratories and quantified here at both the mechanism and the output, that the Allstruder aims to serve as shared infrastructure for the next generation of syringe-based printing.
None.
Thomas Hinton: Writing – review and editing, Writing – original draft, Visualization, Validation, Resources, Project administration, Methodology, Investigation, Formal analysis, Data curation, Conceptualization. Riley Patten: Writing – original draft, Visualization, Validation, Methodology, Investigation, Formal analysis, Data curation, Conceptualization. Antonio J. PereiraTavares: Writing – review and editing, Visualization, Investigation, Formal analysis, Data curation. Cody Crosby: Writing – review and editing, Validation, Investigation, Data curation. Daniel J. Shiwarski: Writing – review and editing, Writing – original draft, Visualization, Validation, Resources, Project administration, Methodology, Investigation, Formal analysis, Data curation, Funding support.
The authors declare the following financial interests and personal relationships which may be considered as potential competing interests: T.J.H. and D.J.S. have an equity stake in FluidForm Bio, Inc, which is a startup company commercializing FRESH 3D printing. No other authors declare that they have known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
We thank members of the Kaplan, Crosby, and Shiwarski laboratories for input and support during the completion of this work and preparation of the manuscript. We thank Vincent Fitzpatrick, Karl Schneider, Enric Casanova-Batlle, Charlie McLean, Tony Tam, Mark Skylar-Scott, Liam Power, Brian Coffin, Jonathan Wright, Martin Sahul, Evelyn Shen, Noé González Rodríguez, Amanda Mejia, Joseph Dorsey, Matthew Griffin, and members of the NYU ITP Biofabrication Class for their efforts to test, adopt, and provide feedback on the Allstruder.
This work was supported by the National Institutes of Health [P41EB027062, and R00HL155777].
During the preparation of this work the authors used Claude and ChatGPT to condense wording and fix grammatical errors. After using this tool or service, the authors reviewed and edited the content as needed and take full responsibility for the content of the published article.
Supplementary data to this article can be found online at doi.org/10.1016/j.ohx.2026.e00827. Data within the supplementary files include CAD files, assembly instructions, slicing profiles, printer mounting hardware, and additional resources. All other information can be found at the Allstruder GitHub page and Zenodo deposit.
Reproduced from Hinton T, Patten R, PereiraTavares AJ, Crosby C, Shiwarski DJ. The Allstruder: an open syringe extruder for every 3D printer. HardwareX 28 (2026) e00827, under the terms of the CC BY 4.0 license. Figures and supplementary data are available with the original article.